Two ring artifact correction methods for computed tomagraphy

  • Beihang University

科研成果: 期刊稿件文章同行评审

摘要

Ring artifacts in computed tomography (CT) images are introduced by detector element's response nonuniformity. Ring artifact degrades the quality of CT image, influences the determination of defect, and hampers post processing and prohibiting quantitative analysis. Two algorithms for ring artifact correction are proposed based on feature recognition and B-form of the spline curve fitting. The former is for the situation that the response of a single detector element or few continuous detector elements is obviously nonuniform. According to the algorithm, the position of the detector element whose response is nonuniform is determined. And a local median filter is executed at the position to depress the vertical streaking in the sinogram. The latter is for the situation that many continuous detector elements' response has small nonuniformity. In this method, the normalized gain factors of the detector elements are gotten through B-form of the spline curve fitting. So detector element's response nonuniformity can be corrected. The two algorithms have their own characteristics which supply each other and compose together a complete ring artifact correction system. The feasibility of the two methods is validated by the experimental data.