TY - JOUR
T1 - The blood heat exchanger in intra-arterial selective cooling infusion for acute ischemic stroke
T2 - A computational fluid-thermodynamics performance, experimental assessment and evaluation on the brain temperature
AU - Jiang, Miaowen
AU - Gao, Yuan
AU - Wu, Chuanjie
AU - Wu, Longfei
AU - Tang, Shoucheng
AU - Yin, Zhichen
AU - Li, Ang
AU - Wang, Kun
AU - Zheng, Shiqiang
AU - Lee, Hangil
AU - Ding, Yuchuan
AU - Li, Ming
AU - Ji, Xunming
N1 - Publisher Copyright:
© 2022 Elsevier Ltd
PY - 2022/6
Y1 - 2022/6
N2 - Intra-arterial selective cooling infusion with the autologous blood (IA-SCAI) is a promising therapeutic hypothermia induction method for conferring neuroprotection to acute ischemic stroke (AIS) patients. The blood heat exchanger (BHE) plays a crucial role in IA-SCAI's cooling capacity. However, there are no BHEs currently available that are specifically designed for the IA-SCAI, which requires a low blood flow to be compatible with cerebral hemodynamics. In an effort to develop a BHE for AIS patients, a prototype of a commercial BHE, Medtronic MYOtherm XP®, was mathematically modeled; specifically, computational fluid dynamics (CFD) was used to analyze its hemo- and thermo-dynamic characteristics under low blood flow including temperature distribution, velocity field and shear stress. Our numerical model predicted the hemolysis index to be 0.0041%–0.0581% inside the BHE with blood flows rates of 10 ml min−1–50 ml min−1. The in vitro heat transfer experiment showed that the BHE efficiently cooled the simulated blood from the initial 37 °C–5.8 °C within 150 s by using cold water (200 ml·min-1, 0 °C). The cooled simulated blood was able to cool the simulated blood in the middle cerebral artery of an artificial circulating system from 37 °C to 16.8 °C–33.7 °C depending on the blood perfusion rate (10–50 ml/min). A biological heat transfer mathematical model showed that brain tissue could be cooled by 2 °C within the initial 1min of infusion. This study verified the feasibility of using a commercial BHE for IA-SCAI and provided insights into its cooling capacity for therapeutic hypothermia.
AB - Intra-arterial selective cooling infusion with the autologous blood (IA-SCAI) is a promising therapeutic hypothermia induction method for conferring neuroprotection to acute ischemic stroke (AIS) patients. The blood heat exchanger (BHE) plays a crucial role in IA-SCAI's cooling capacity. However, there are no BHEs currently available that are specifically designed for the IA-SCAI, which requires a low blood flow to be compatible with cerebral hemodynamics. In an effort to develop a BHE for AIS patients, a prototype of a commercial BHE, Medtronic MYOtherm XP®, was mathematically modeled; specifically, computational fluid dynamics (CFD) was used to analyze its hemo- and thermo-dynamic characteristics under low blood flow including temperature distribution, velocity field and shear stress. Our numerical model predicted the hemolysis index to be 0.0041%–0.0581% inside the BHE with blood flows rates of 10 ml min−1–50 ml min−1. The in vitro heat transfer experiment showed that the BHE efficiently cooled the simulated blood from the initial 37 °C–5.8 °C within 150 s by using cold water (200 ml·min-1, 0 °C). The cooled simulated blood was able to cool the simulated blood in the middle cerebral artery of an artificial circulating system from 37 °C to 16.8 °C–33.7 °C depending on the blood perfusion rate (10–50 ml/min). A biological heat transfer mathematical model showed that brain tissue could be cooled by 2 °C within the initial 1min of infusion. This study verified the feasibility of using a commercial BHE for IA-SCAI and provided insights into its cooling capacity for therapeutic hypothermia.
KW - Blood heat exchanger
KW - Hemolysis
KW - Numerical simulation
KW - Stroke
KW - Therapeutic hypothermia
UR - https://www.scopus.com/pages/publications/85127525289
U2 - 10.1016/j.compbiomed.2022.105497
DO - 10.1016/j.compbiomed.2022.105497
M3 - 文章
C2 - 35398811
AN - SCOPUS:85127525289
SN - 0010-4825
VL - 145
JO - Computers in Biology and Medicine
JF - Computers in Biology and Medicine
M1 - 105497
ER -