跳到主要导航 跳到搜索 跳到主要内容

Systematic studies of scintillation detector with timing resolution of 10 ps for heavy ion beam

  • S. Nishimura*
  • , M. Kurata-Nishimura
  • , K. Morimoto
  • , Y. Nishi
  • , A. Ozawa
  • , T. Yamaguchi
  • , T. Ohnishi
  • , T. Zheng
  • , M. Chiba
  • , I. Tanihata
  • *此作品的通讯作者
  • RIKEN
  • Peking University

科研成果: 期刊稿件文章同行评审

摘要

We have tested Time-Of-Flight detectors consisting of several combinations of scintillators and photomultiplier tubes for the identification of produced heavy ion beam. Timing resolution of 6.8 ps (RMS) is achieved with energy deposited of 150-160 MeV in the plastic scintillator, where 40Ar beam is used as a primary beam at the total energies of 3.8 GeV. Our systematic studies suggest the importance of maximizing the number of photoelectrons, where the timing resolution seems to be scaled by the pulse height according to the characteristic function of the photomultiplier tube. In addition, scintillator with faster timing property seems to be also essential for better timing resolution.

源语言英语
页(从-至)377-388
页数12
期刊Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
510
3
DOI
出版状态已出版 - 11 9月 2003
已对外发布

学术指纹

探究 'Systematic studies of scintillation detector with timing resolution of 10 ps for heavy ion beam' 的科研主题。它们共同构成独一无二的学术指纹。

引用此