跳到主要导航 跳到搜索 跳到主要内容

Suppression of vapor cell temperature error for spin-exchange-relaxation-free magnetometer

科研成果: 期刊稿件文献综述同行评审

摘要

This paper presents a method to reduce the vapor cell temperature error of the spin-exchange-relaxation-free (SERF) magnetometer. The fluctuation of cell temperature can induce variations of the optical rotation angle, resulting in a scale factor error of the SERF magnetometer. In order to suppress this error, we employ the variation of the probe beam absorption to offset the variation of the optical rotation angle. The theoretical discussion of our method indicates that the scale factor error introduced by the fluctuation of the cell temperature could be suppressed by setting the optical depth close to one. In our experiment, we adjust the probe frequency to obtain various optical depths and then measure the variation of scale factor with respect to the corresponding cell temperature changes. Our experimental results show a good agreement with our theoretical analysis. Under our experimental condition, the error has been reduced significantly compared with those when the probe wavelength is adjusted to maximize the probe signal. The cost of this method is the reduction of the scale factor of the magnetometer. However, according to our analysis, it only has minor effect on the sensitivity under proper operating parameters.

源语言英语
期刊论文编号083103
期刊Review of Scientific Instruments
86
8
DOI
出版状态已出版 - 1 8月 2015

学术指纹

探究 'Suppression of vapor cell temperature error for spin-exchange-relaxation-free magnetometer' 的科研主题。它们共同构成独一无二的学术指纹。

引用此