跳到主要导航 跳到搜索 跳到主要内容

Suppression of the magnetic noise response caused by elliptically polarized light in an optical rotation detection system

  • Beihang University
  • Zhejiang Lab

科研成果: 期刊稿件文章同行评审

摘要

We analyze and suppress the magnetic noise response in optical rotation detection system (ORDS) in atomic magnetometers in this study. Because of the imperfections of the optical elements, the probe light is actually elliptically polarized in ORDS, which can polarize the atom ensemble and cause the responses to the three-axis magnetic noise. We theoretically analyze the frequency responses to the magnetic noise, and prove that the responses are closely associated with the DC magnetic field. The values of the DC magnetic fields are calculated with special frequency points, called ‘break points’, in the transverse responses. We reveal the relationships between the DC magnetic field and the sensitivities of ORDS, and effectively suppress the magnetic noise responses with the residual magnetic field compensation. Finally, the sensitivity of ORDS is improved by approximately two times at 10-20 Hz.

源语言英语
页(从-至)3854-3865
页数12
期刊Optics Express
30
3
DOI
出版状态已出版 - 31 1月 2022

学术指纹

探究 'Suppression of the magnetic noise response caused by elliptically polarized light in an optical rotation detection system' 的科研主题。它们共同构成独一无二的学术指纹。

引用此