TY - JOUR
T1 - Sequential Scan-Based High-Resolution 3-D Open-Sided Magnetic Particle Imaging System
AU - He, Jie
AU - Zhang, Haoran
AU - Lei, Siao
AU - Li, Guanghui
AU - Li, Yimeng
AU - Zhu, Tao
AU - Qian, Zhumei
AU - Xiong, Fei
AU - Feng, Yuan
AU - An, Yu
AU - Tian, Jie
N1 - Publisher Copyright:
© 1963-2012 IEEE.
PY - 2024
Y1 - 2024
N2 - Magnetic particle imaging (MPI) is an emerging molecular imaging technique that relies on the nonlinear magnetization response of superparamagnetic iron oxide nanoparticles. Open-sided MPI (OS-MPI) has attracted considerable interest for its potential to enhance interventional procedures. However, improving the spatial resolution of OS-MPI poses a significant challenge, mainly due to the inefficiencies of selection coils in generating, rotating, and shifting a high-gradient field-free line (FFL). To address this issue, we propose a novel approach called the permanent magnet-based sequential scan to develop a high-resolution 3-D OS-MPI system with ultralow power consumption. By utilizing permanent magnets, we can create high-gradient selection fields without any power consumption. The sequential scan technique involves rotating the FFL angle by angle and shifting the FFL layer by layer. To maintain thermal stability during sequential scans, we used drive coils wound with hollow copper wires and implemented forced oil cooling. Additionally, measures such as copper shielding, feed-through compensation, and filtering were employed to minimize interference from the permanent magnets, drive coils, and motor drivers. Experimental results demonstrated that the developed system successfully achieved a spatial resolution of 1 mm while operating with an overall low power consumption of 282.16 W.
AB - Magnetic particle imaging (MPI) is an emerging molecular imaging technique that relies on the nonlinear magnetization response of superparamagnetic iron oxide nanoparticles. Open-sided MPI (OS-MPI) has attracted considerable interest for its potential to enhance interventional procedures. However, improving the spatial resolution of OS-MPI poses a significant challenge, mainly due to the inefficiencies of selection coils in generating, rotating, and shifting a high-gradient field-free line (FFL). To address this issue, we propose a novel approach called the permanent magnet-based sequential scan to develop a high-resolution 3-D OS-MPI system with ultralow power consumption. By utilizing permanent magnets, we can create high-gradient selection fields without any power consumption. The sequential scan technique involves rotating the FFL angle by angle and shifting the FFL layer by layer. To maintain thermal stability during sequential scans, we used drive coils wound with hollow copper wires and implemented forced oil cooling. Additionally, measures such as copper shielding, feed-through compensation, and filtering were employed to minimize interference from the permanent magnets, drive coils, and motor drivers. Experimental results demonstrated that the developed system successfully achieved a spatial resolution of 1 mm while operating with an overall low power consumption of 282.16 W.
KW - Field-free line (FFL)
KW - magnetic particle imaging (MPI)
KW - open-sided
KW - sequential scan
UR - https://www.scopus.com/pages/publications/85192963594
U2 - 10.1109/TIM.2024.3396830
DO - 10.1109/TIM.2024.3396830
M3 - 文章
AN - SCOPUS:85192963594
SN - 0018-9456
VL - 73
SP - 1
EP - 12
JO - IEEE Transactions on Instrumentation and Measurement
JF - IEEE Transactions on Instrumentation and Measurement
M1 - 4007112