Rapid calibration of 3D freehand ultrasound for vessel intervention

  • K. Luan*
  • , H. Liao
  • , T. Ohya
  • , J. Wang
  • , I. Sakuma
  • *此作品的通讯作者
  • The University of Tokyo

科研成果: 书/报告/会议事项章节会议稿件同行评审

摘要

To create a freehand three-dimensional (3D) ultrasound (US) system for image-guided vessel interventions, a US probe calibration must be performed to build 3D volumes from 2D vessel images. In this paper, a novel calibration method of 3D freehand US is described. It simply images the tip of a tracked needle that could freely move around within the coupling medium while the probe is fixed. The physical coordinates of the tip and the probe location are simultaneously recorded. Image coordinates of the tip are mapped to physical points. A RANSAC algorithm is applied to minimize the error matching image points and physical points. An accuracy evaluation experiment is performed to verify the proposed method. Results show that the needle tip can be accurately located and the RMS calibration error is 1.56mm. A phantom experiment and an in vivo experiment verify the availability of the technique in clinic. That means the present method is applicable to the calibration of a 3D freehand US in vessel interventions.

源语言英语
主期刊名5th Kuala Lumpur International Conference on Biomedical Engineering 2011, BIOMED 2011
716-719
页数4
DOI
出版状态已出版 - 2011
活动5th Kuala Lumpur International Conference on Biomedical Engineering, BIOMED 2011, Held in Conjunction with the 8th Asian Pacific Conference on Medical and Biological Engineering, APCMBE 2011 - Kuala Lumpur, 马来西亚
期限: 20 6月 201123 6月 2011

出版系列

姓名IFMBE Proceedings
35 IFMBE
ISSN(印刷版)1680-0737

会议

会议5th Kuala Lumpur International Conference on Biomedical Engineering, BIOMED 2011, Held in Conjunction with the 8th Asian Pacific Conference on Medical and Biological Engineering, APCMBE 2011
国家/地区马来西亚
Kuala Lumpur
时期20/06/1123/06/11

学术指纹

探究 'Rapid calibration of 3D freehand ultrasound for vessel intervention' 的科研主题。它们共同构成独一无二的学术指纹。

引用此