摘要
We explore the impact of pumping beams with different transverse intensity profiles on the performance of the spin-exchange relaxation-free (SERF) atomic magnetometers (AMs). We conduct experiments comparing the traditional Gaussian optically-pumped AM with that utilizing the flat-top optically-pumped (FTOP) method. Our findings reveal that the FTOP-based approach outperforms the conventional method, exhibiting a larger response, a narrower magnetic resonance linewidth, and a superior low-frequency noise performance. Specifically, the use of FTOP method leads to a 16% enhancement in average sensitivity within 1 Hz-30 Hz frequency range. Our research emphasizes the significance of achieving transverse polarization uniformity in AMs, providing insights for future optimization efforts and sensitivity improvements in miniaturized magnetometers.
| 源语言 | 英语 |
|---|---|
| 文章编号 | 060703 |
| 期刊 | Chinese Physics B |
| 卷 | 33 |
| 期 | 6 |
| DOI | |
| 出版状态 | 已出版 - 1 5月 2024 |