跳到主要导航 跳到搜索 跳到主要内容

Optical system for microscopic observation and strain measurement at high temperature

  • The University of Tokyo

科研成果: 期刊稿件文章同行评审

摘要

A system of optical microscopy has been developed for high-temperature direct observation and strain measurement. Observations at high temperatures are carried out using a UV CCD camera and UV illumination while thermal radiation is cut off by a UV light transmission filter. Observation of a Vickers indentation on polycrystalline -Al2O3 revealed that sufficient image contrast was maintained up to 1400 °C. The potential of the proposed system for strain measurement via digital image correlation was evaluated through measuring the thermal strain of polycrystalline -Al 2O3; the obtained coefficient of thermal expansion agreed well with the reported values.

源语言英语
文章编号025002
期刊Measurement Science and Technology
25
2
DOI
出版状态已出版 - 1 2月 2014