Method to quickly estimate T1 value by suppressing spin exchange relaxation and magnetic field gradient relaxation in atomic sensors

  • Beihang University
  • Hefei National Laboratory
  • Hangzhou Institute of Extremely-Weak Magnetic Field Major National Science and Technology Infrastructure

科研成果: 期刊稿件文章同行评审

摘要

The accuracy of atomic sensor is determined by the transverse nuclear spin relaxation time (T2). As the longitudinal nuclear spin relaxation time (T1) is the upper limit of T2, it becomes an important metric for evaluating the merits of alkali vapor cells. However, the conventional method for measuring T1 of pulse delay takes dozens of minutes, which will increase the effect of random errors during the testing process. In order to solve this problem, a method to quickly estimate the value of T1 is proposed by suppressing the influence of spin exchange relaxation and magnetic field gradient relaxation terms on T2. Experimental results indicate that the accuracy is maintained at above 95%, and the testing time is reduced to less than 5%.

源语言英语
页(从-至)15380-15389
页数10
期刊Optics Express
32
9
DOI
出版状态已出版 - 22 4月 2024

指纹

探究 'Method to quickly estimate T1 value by suppressing spin exchange relaxation and magnetic field gradient relaxation in atomic sensors' 的科研主题。它们共同构成独一无二的指纹。

引用此