TY - JOUR
T1 - Magnetic resonance linewidths independent with the laser linewidth in coherent population trapping (CPT) magnetometers
AU - Du, Pengcheng
AU - Yang, Feifan
AU - Zhou, Yuansheng
AU - Quan, Wei
AU - Li, Jin
N1 - Publisher Copyright:
© 2024
PY - 2025/1/30
Y1 - 2025/1/30
N2 - Coherent population trapping (CPT) magnetometers are the magnetic measuring instrument that utilize the interference effect between light and atoms. It has a broad range of applications in weak magnetic measurement. However, current CPT magnetometers have low sensitivity, which does not satisfy the ultra-high sensitivity requirements of weak magnetic measurement. This paper is aimed to improve this problem by studying the relationship between the sensitivity of CPT magnetometer and the laser linewidth. It has been concluded that the magnetic resonance linewidth of the CPT system remains independent of the laser linewidth according to the Liouville equation, firstly. Furthermore, the correctness of this theory has been verified through design experiments. Additionally, we have identified certain issues that require attention in the miniaturization and chip-scale CPT magnetometer on vertical cavity surface emitting laser (VCSEL).
AB - Coherent population trapping (CPT) magnetometers are the magnetic measuring instrument that utilize the interference effect between light and atoms. It has a broad range of applications in weak magnetic measurement. However, current CPT magnetometers have low sensitivity, which does not satisfy the ultra-high sensitivity requirements of weak magnetic measurement. This paper is aimed to improve this problem by studying the relationship between the sensitivity of CPT magnetometer and the laser linewidth. It has been concluded that the magnetic resonance linewidth of the CPT system remains independent of the laser linewidth according to the Liouville equation, firstly. Furthermore, the correctness of this theory has been verified through design experiments. Additionally, we have identified certain issues that require attention in the miniaturization and chip-scale CPT magnetometer on vertical cavity surface emitting laser (VCSEL).
KW - Coherent population trapping (CPT) magnetometers
KW - Laser linewidth
KW - Magnetic resonance linewidths
UR - https://www.scopus.com/pages/publications/85203010584
U2 - 10.1016/j.measurement.2024.115654
DO - 10.1016/j.measurement.2024.115654
M3 - 文章
AN - SCOPUS:85203010584
SN - 0263-2241
VL - 240
JO - Measurement: Journal of the International Measurement Confederation
JF - Measurement: Journal of the International Measurement Confederation
M1 - 115654
ER -