跳到主要导航 跳到搜索 跳到主要内容

Integration of microsensor for microsurgery robot's end-effector

  • Beihang University
  • Capital Medical University

科研成果: 期刊稿件文章同行评审

摘要

To enhance the effect of robotic microsurgery, the microsensors are integrated on the robot's end-effector. On the basis of the requirements presented for the integration design, measuring mechanism for the robotic end trephine's force and cutting depth are studied. Force microsensor and position microsensor are used to measure surgical information of the force and depth. Measuring mechanism was achieved by means of linear sliding bearing and differential measuring structure. The sensor data board was developed. With the power spectral estimation of sensor data, two digital filtering methods are proposed, to help eliminate the interference to the original microsensor signal. They are the filtering method of lowpass-bandstop serial structure suitable for a PC, and a shift average filtering algorithm suitable for the sensor data board, respectively. The experimental results show that the integration of microsensors for microsurgery robot's end-effector can satisfy the design requirements, and the robotic end trephine can accurately fulfill the surgical task of corneal cutting.

源语言英语
页(从-至)205-209
页数5
期刊Frontiers of Mechanical Engineering in China
2
2
DOI
出版状态已出版 - 4月 2007

学术指纹

探究 'Integration of microsensor for microsurgery robot's end-effector' 的科研主题。它们共同构成独一无二的学术指纹。

引用此