摘要
A laser frequency stabilization technique based on Pound -Drever-Hall (PDH) method was introduced and analyzed in detail. The influences of the Fabry-Perot(F-P) mirror's reflectivity and loss of F-P cavity were analyzed by numerical simulation. The theoretical linear dynamic range (LDR) and sensitivity were 9.6 MHz and 0.2 MHz-1 without loss respectively if the reflectivity was 0.2. Meanwhile, the LDR was less than 14.5 MHz and the sensitivity preceded 0.05 MHz-1 in a situation considering that the cavity loss was about 1%. The results demonstrate that both the two factors cause significant influence on the LDR and sensitivity of the system.
| 源语言 | 英语 |
|---|---|
| 页(从-至) | 3699-3701 |
| 页数 | 3 |
| 期刊 | Infrared and Laser Engineering |
| 卷 | 43 |
| 期 | 11 |
| 出版状态 | 已出版 - 25 11月 2014 |
指纹