In vitro analysis of wearing of hip joint prostheses composed of different contact materials

  • Jian Su
  • , Jian Jun Wang
  • , Shi Tong Yan
  • , Min Zhang
  • , Hui Zhi Wang
  • , Ning Ze Zhang
  • , Yi Chao Luan
  • , Cheng Kung Cheng*
  • *此作品的通讯作者
  • Beihang University
  • Beijing Institute of Medical Device Testing
  • Shanghai Jiao Tong University

科研成果: 期刊稿件文章同行评审

摘要

Cobalt-chromium-molybdenum alloy (CoCrMo) and ceramic are the two most common materials for the femoral head in hip joint prostheses, and the acetabular liner is typically made from ultra-high molecular weight polyethylene (UHMWPE), highly cross-linked polyethylene (XLPE), or highly cross-linked polyethylene blended with Vitamin E (VEXLPE). The selection of suitable materials should consider both wear performance and cost-effectiveness. This study compared the wear rate between different friction pairs using a hip joint simulator and then rec-ommended a suitable prosthesis based on the corresponding processing technology and cost. All wear simulations were performed in accordance with ISO 14242, using the same hip joint simulator and same test conditions. This study found that when using the same material for the femoral head, the XLPE and VEXLPE liners had a lower wear rate than the UHMWPE liners, and the wear rate of the XLPE liners increased after blending with Vitamin E (VEXLPE). There was no significant difference in the wear rate of XLPE when using a CoCrMo or ceramic head. Considering the wear rate and cost-effectiveness, a CoCrMo femoral head with an accompanying XLPE liner is recom-mended as the more suitable combination for hip prostheses.

源语言英语
文章编号3805
期刊Materials
14
14
DOI
出版状态已出版 - 2 7月 2021

指纹

探究 'In vitro analysis of wearing of hip joint prostheses composed of different contact materials' 的科研主题。它们共同构成独一无二的指纹。

引用此