@inproceedings{712d73e3998e4878aab7171cfa63f6ed,
title = "High sensitivity sensing membranes of high addressable potentiometric sensor for bio material detection",
abstract = "In this study, for detection Bio materials, a high sensitivity Light Addressable Potentiometric sensor (LAPS) device which used the Si3N4/SiO2/Si, Ta2O5/SiO2/Si structure were fabricated, respectively. The Penicillinase was immobilized on the sensing membranes using the self-assembled monolayer (SAM) method. The LabVIEW and GPIB were used for detecting the enzyme reaction. After that, according to the detected result, the sensitivity of the Si3N4 sensing membrane LAPS device is known as 60mV/decade and the Ta2O5 sensing membrane LAPS device is 74 mV/decade in the range of 0.1mM\textasciitilde{}10mM of the penicillin{\textquoteright}s concentration.",
keywords = "Bio material, LAPS, TaO",
author = "Jang, \{Su Won\} and Heng Yuan and Kim, \{Do Eok\} and Kim, \{Kyu Jin\} and Kang, \{Byoung Ho\} and Eum, \{Nyeon Sik\} and Kang, \{Shin Won\}",
note = "Publisher Copyright: {\textcopyright} International Federation for Medical and Biological Engineering 2007.; 10th World Congress on Medical Physics and Biomedical Engineering, WC 2006 ; Conference date: 27-08-2006 Through 01-09-2006",
year = "2007",
doi = "10.1007/978-3-540-36841-0\_329",
language = "英语",
isbn = "9783540368397",
series = "IFMBE Proceedings",
publisher = "Springer Verlag",
number = "1",
pages = "1349--1352",
editor = "Kim, \{Sun I.\} and Suh, \{Tae Suk\}",
booktitle = "IFMBE Proceedings",
address = "德国",
edition = "1",
}