跳到主要导航 跳到搜索 跳到主要内容

Hemoglobin-titanate composite based biosensor for the amperometric determination of hydrogen peroxide in acidic medium

  • Yihua Wu*
  • , Qiuming Gao
  • , Linpu Shi
  • , Lu Gao
  • *此作品的通讯作者
  • CAS - Shanghai Institute of Ceramics

科研成果: 期刊稿件文章同行评审

摘要

A hemoglobin-titanate composite based biosensor was chosen for determination of H2O2 in an acidic medium. CV results of the Hb-titanate modified pyrolytic graphite electrode showed a pair of well-defined, quasi-reversible redox peaks centered at - 246 mV (vs. Ag/AgCl) in a pH 5.0 HAc-NaAc buffer solution. The modified electrode exhibited good electrocatalytic response for monitoring H2O2 and had a large linear detection range from 20 μM to 3.2 mM with a detection limit of 8 μM (S/N = 3) and a sensitivity of 29.7 mA M-1 cm-2 in the pH 5.0 solution. The biosensor also possessed good long term storage stability.

源语言英语
页(从-至)904-908
页数5
期刊Electroanalysis
21
8
DOI
出版状态已出版 - 4月 2009
已对外发布

学术指纹

探究 'Hemoglobin-titanate composite based biosensor for the amperometric determination of hydrogen peroxide in acidic medium' 的科研主题。它们共同构成独一无二的学术指纹。

引用此