Gradient phase and amplitude errors in atomic magnetic gradiometers for biomagnetic imaging systems

  • Beihang University
  • Hefei National Laboratory

科研成果: 期刊稿件文章同行评审

摘要

The cross-axis projection error (CAPE) caused by residual magnetic fields has recently attracted widespread attention. In this study, we propose a more specific theoretical model and expand the CAPE in gradient measurements. We first report that differences in relaxation rate and residual magnetic field between optically pumped magnetometers (OPMs) introduce a significant error term in the output of OPM gradiometers, referred to as the gradient phase error. Furthermore, when the longitudinal field compensation is inadequate, the interaxial response interference of a single OPM is prominent, resulting in an amplitude distortion of the signal. This is further amplified in the gradiometer configuration, introducing the gradient amplitude error. Our experiments demonstrated that the efficacy of mitigating common-mode noise of OPM gradiometers was significantly impaired when existing the gradient errors. In addition, a simulation with a magnetoencephalography (MEG) system illustrated an induced source localization error of exceeding 2 cm, severely compromising the localization accuracy of OPM-MEG systems.

源语言英语
文章编号109250
期刊iScience
27
3
DOI
出版状态已出版 - 15 3月 2024

指纹

探究 'Gradient phase and amplitude errors in atomic magnetic gradiometers for biomagnetic imaging systems' 的科研主题。它们共同构成独一无二的指纹。

引用此