Fast calibration methods using cosmic rays for a neutron detection array

  • Zai Hong Yang*
  • , Yan Lin Ye
  • , Jun Xiao
  • , Hai Bo You
  • , Hong Na Liu
  • , Ye Lei Sun
  • , Zi Heng Wang
  • , Jie Chen
  • *此作品的通讯作者
  • Peking University

科研成果: 期刊稿件文章同行评审

摘要

An overall irradiation and calibration technique was introduced and applied to a test scintillation detector array. An integral conversion method was used to reduce the nonlinearity of the time difference spectrum, and to improve the position determination especially for positions close to the two ends of a long scintillation bar. An overall position resolution of about 3.0 cm (FWHM) was extracted from the residual analysis method and verified by a direct measurement. Energy calibration was also realized by selecting cosmic rays at different incident angles. The bulk light attenuation lengths for the four test bars were also determined. It is demonstrated that these methods are especially efficient for calibrating large and complex detector arrays.

源语言英语
页(从-至)222-227
页数6