摘要
Neutron detection efficiency for a long wedge-shaped plastic scintillation detector was measured by using the Am/Be neutron source. The overall efficiency attains 23% at around 2 MeV and decreases to 15% at about 5 MeV. The experimental results are in agreement with the Monte Carlo simulation calculations.
| 源语言 | 英语 |
|---|---|
| 页(从-至) | 473-477 |
| 页数 | 5 |
| 期刊 | IEEE Transactions on Nuclear Science |
| 卷 | 52 |
| 期 | 1 III |
| DOI | |
| 出版状态 | 已出版 - 2月 2005 |
| 已对外发布 | 是 |