Effects of reverse deployment of cone-shaped vena cava filter on improvements in hemodynamic performance in vena cava

  • Ying Chen*
  • , Zaipin Xu*
  • , Xiaoyan Deng*
  • , Shibo Yang
  • , Wenchang Tan
  • , Yubo Fan
  • , Yong Han
  • , Yubin Xing
  • *此作品的通讯作者
  • Beijing Institute of Economics and Management
  • Peking University
  • Beihang University
  • Guizhou University
  • Sichuan University of Science & Engineering
  • Guizhou Institute of Animal Husbandry and Veterinary Science
  • General Hospital of People's Liberation Army

科研成果: 期刊稿件文章同行评审

摘要

Background: Cone-shaped vena cava filters (VCFs) are widely used to treat venous thromboembolism. However, in the long term, the problem of occlusion persists even after the filter is deployed. A previous study hypothesized that the reverse deployment of a cone-shaped VCFs may prevent filter blockage. Methods: To explore this hypothesis, a comparative study of the traditional and reverse deployments of VCFs was conducted using a computational fluid dynamics approach. The distribution of wall shear stress (WSS) and shear stress-related parameters were calculated to evaluate the differences in hemodynamic effects between both conditions. In the animal experiment, we reversely deployed a filter in the vena cava of a goat and analyzed the blood clot distribution in the filter. Results: The numerical simulation showed that the reverse deployment of a VCF resulted in a slightly higher shear rate on the thrombus, and no reductions in the oscillating shear index (OSI) and relative residence time (RRT) on the vessel wall. Comparing the traditional method with the reversely deployed cases, the shear rate values is 16.49 and 16.48 1/s, respectively; the minimal OSI values are 0.01 and 0.04, respectively; in the vicinity of the VCF, the RRT values are both approximately 5 1/Pa; and the WSS is approximately 0.3 Pa for both cases. Therefore, the reverse deployment of cone-shaped filters is not advantageous when compared with the traditional method in terms of local hemodynamics. However, it is effective in capturing thrombi in the short term, as demonstrated via animal experiments. The reversely deployed cone-shaped filter captured the thrombi at its center in the experiments. Conclusions: Thus, the reverse deployment of cone-shaped filters is not advantageous when compared with the traditional method in terms of local hemodynamics. Therefore, we would not suggest the reverse deployment of the cone-shaped filter in the vena cava to prevent a potentially fatal pulmonary embolism.

源语言英语
期刊论文编号19
期刊BioMedical Engineering Online
20
1
DOI
出版状态已出版 - 12月 2021

学术指纹

探究 'Effects of reverse deployment of cone-shaped vena cava filter on improvements in hemodynamic performance in vena cava' 的科研主题。它们共同构成独一无二的学术指纹。

引用此