跳到主要导航 跳到搜索 跳到主要内容

Effects of pump laser intensity on the cell temperature working point in a K-Rb-21 Ne spin-exchange relaxation-free co-magnetometer

  • Beihang University
  • Beihang Hangzhou Innovation Institute

科研成果: 期刊稿件 › 文章 › 同行评审

摘要

The cell temperature working point optimization of the spin-exchange relaxation-free (SERF) co-magnetometer is studied theoretically and experimentally in this article. Based on the steady-state solution of the Bloch equations, the steady-state response model of the K-Rb-21Ne SERF co-magnetometer output signal with cell temperature is established in this paper. And combined with the model, a method to find the optimal working point of the cell temperature that incorporates the pump laser intensity is proposed. The scale factor of the co-magnetometer under different pump laser intensities and cell temperatures is obtained experimentally, and the long-term stability of the co-magnetometer at the different cell temperatures with corresponding pump laser intensities is measured. The results show that the bias instability of the co-magnetometer is reduced from 0.0311 deg/h to 0.0169 deg/h by obtaining the optimal working point of the cell temperature, which verifies the validity and accuracy of the theoretical derivation and the proposed method.

源语言英语
页(从-至)8342-8351
页数10
期刊Optics Express
卷31
期5
DOI
出版状态已出版 - 27 2月 2023

学术指纹

探究 'Effects of pump laser intensity on the cell temperature working point in a K-Rb-21 Ne spin-exchange relaxation-free co-magnetometer' 的科研主题。它们共同构成独一无二的学术指纹。

引用此