Dew point measurements using montmorillonite (MTT) and molybdenum disulfide (MoS2) coated QCM sensors

  • Ning Li
  • , Qiyang Huang
  • , Jing Nie*
  • , Xiaofeng Meng
  • , Zhongying Wang
  • , Yichuan Wu
  • , Mingjing Qi
  • , Zhiwei Liu
  • , Baoxia Mi
  • , Liwei Lin
  • *此作品的通讯作者
  • Beihang University
  • University of California at Berkeley
  • Tsinghua University

科研成果: 期刊稿件文章同行评审

摘要

Due to the influence of temperature, the traditional quartz crystal microbalance (QCM) dew point sensors require an independent study on the temperature frequency characteristics (TFC) for calibrations as the frequency response is susceptible to external interferences. QCM dew point sensors with either montmorillonite (MTT) or molybdenum disulfide (MoS2) coatings together with a dew point recognition algorithm are demonstrated for dew point recognitions. The frequency signals output from three kinds of QCM sensors are processed by the proposed recognition algorithm and compared with the standard dew point data provided by MICHELL S4000. The results show QCM sensors with MTT coating have the best accuracy and repeatability test for 5 days in range of 3 ℃ DP–15 ℃ DP environment has shown the maximum relative error as compared with the standard dew-point values is less than ±0.3 ℃ DP. As such, this work shows the feasibility of utilizing QCM humidity sensors without the influence of temperature on frequency measurements with good repeatability.

源语言英语
页(从-至)122-129
页数8
期刊Sensors and Actuators B: Chemical
279
DOI
出版状态已出版 - 15 1月 2019

指纹

探究 'Dew point measurements using montmorillonite (MTT) and molybdenum disulfide (MoS2) coated QCM sensors' 的科研主题。它们共同构成独一无二的指纹。

引用此