TY - JOUR
T1 - Design of a miniaturized forearm magnetomyography detection system based on biplanar coils
AU - Yuan, Shuai
AU - Li, Haijun
AU - Yang, Fan
AU - Zhang, Leran
AU - Ma, Yuzheng
AU - Wang, Zheng
AU - Qi, Yuankun
AU - Shi, Minxia
N1 - Publisher Copyright:
© 2026 IOP Publishing Ltd. All rights, including for text and data mining, AI training, and similar technologies, are reserved. This article is available under the terms of the IOP-Standard License.
PY - 2026/8/7
Y1 - 2026/8/7
N2 - Magnetomyography (MMG) is a precise method for assessing muscle conditions, particularly when measured with high-sensitivity optically pumped magnetometers. However, the MMG signal amplitude, typically on the order of pico-Tesla (pT), is susceptible to interference from the geomagnetic field. Traditional shielding systems, which are effective in mitigating this interference, are nonetheless bulky and occupy considerable space, thereby limiting their portability and hindering their widespread application. In this study, we propose a compensation system for a compact forearm-based MMG detection device utilizing biplanar coils. This system effectively addresses the challenges posed by uneven and excessive residual magnetic fields within the miniaturized device, achieving up to a 91.9% reduction in residual magnetic fields from multiple directions. Using this enhanced device, MMG signals are detected from the forearms of a subject under various grip strengths. Time-domain analyses of these signals provide valuable insights into grip strength.
AB - Magnetomyography (MMG) is a precise method for assessing muscle conditions, particularly when measured with high-sensitivity optically pumped magnetometers. However, the MMG signal amplitude, typically on the order of pico-Tesla (pT), is susceptible to interference from the geomagnetic field. Traditional shielding systems, which are effective in mitigating this interference, are nonetheless bulky and occupy considerable space, thereby limiting their portability and hindering their widespread application. In this study, we propose a compensation system for a compact forearm-based MMG detection device utilizing biplanar coils. This system effectively addresses the challenges posed by uneven and excessive residual magnetic fields within the miniaturized device, achieving up to a 91.9% reduction in residual magnetic fields from multiple directions. Using this enhanced device, MMG signals are detected from the forearms of a subject under various grip strengths. Time-domain analyses of these signals provide valuable insights into grip strength.
KW - biplanar coils
KW - grip strengths
KW - magnetic shielding
KW - magnetomyography (MMG)
KW - optically pumped magnetometers
UR - https://www.scopus.com/pages/publications/105047038411
U2 - 10.1088/1361-6463/ae8f52
DO - 10.1088/1361-6463/ae8f52
M3 - 文章
AN - SCOPUS:105047038411
SN - 0022-3727
VL - 59
JO - Journal of Physics D: Applied Physics
JF - Journal of Physics D: Applied Physics
IS - 31
M1 - 315004
ER -