跳到主要导航 跳到搜索 跳到主要内容

Construction and calibration of the multi-neutron correlation spectrometer at Peking University

  • H. B. You
  • , Z. H. Yang
  • , Y. L. Ye*
  • , Z. H. Li
  • , Y. C. Ge
  • , J. L. Lou
  • , Q. T. Li
  • , J. Xiao
  • , R. Qiao
  • , Z. Y. Tian
  • , Y. L. Sun
  • , H. N. Liu
  • , J. Chen
  • , J. Wu
  • , B. Yang
  • , K. Yang
  • , J. S. Wang
  • , Y. Y. Yang
  • , P. Ma
  • , J. B. Ma
  • S. L. Jin, J. L. Han
*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

摘要

A Multi-Neutron Correlation Spectrometer (MuNCoS) composed of 80 detector modules was constructed at Peking University. A position resolution of less than 3 cm (FWHM) for a typical long scintillation module was obtained based on cosmic ray tests. A calibration method using an unstable nucleus beam of 11Be, together with the fragment identification by using a zero degree telescope, has been successfully developed. Neutron detection efficiencies of 15(1)% and 17(2)% were achieved for neutron energy intervals of 30-40 MeV and 40-60 MeV, respectively, and at a signal charge threshold of 4 MeVee. The experimentally determined neutron detection efficiencies agree well with the simulation based on the GEANT4 code.