摘要
We report on an x-ray cone-beam differential phase-contrast computed laminography (DPC-CL) method for tomographic reconstruction of thin and lamellar objects. We describe the specific scan geometry of DPC-CL, which consists of a Talbot-Lau grating interferometer and a lab-based x-ray tube source, and derive a filtered back-projection (FBP) reconstruction algorithm. The experimental results of a flat sphere phantom and a piece of ham demonstrate the validity of the proposed technique. The existing DPC-CL methods are based on synchrotron sources and the parallel-beam geometry. In contrast, our approach adopts a more accessible x-ray tube source and a cone-beam geometry. Therefore it significantly widens the application range of phase-contrast laminography, particularly in practical laboratory settings, beyond applications at large-scale synchrotron facilities.
| 源语言 | 英语 |
|---|---|
| 期刊论文编号 | 68002 |
| 期刊 | Europhysics Letters |
| 卷 | 106 |
| 期 | 6 |
| DOI | |
| 出版状态 | 已出版 - 1 6月 2014 |
学术指纹
探究 'Cone-beam differential phase-contrast laminography with x-ray tube source' 的科研主题。它们共同构成独一无二的学术指纹。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver