Chromatic aberrations correction for imaging spectrometer based on acousto-optic tunable filter with two transducers

  • Beihang University

科研成果: 期刊稿件文章同行评审

摘要

The acousto-optic tunable filter (AOTF) with wide wavelength range and high spectral resolution has long crystal and two transducers. A longer crystal length leads to a bigger chromatic focal shift and the double-transducer arrangement induces angular mutation in diffracted beam, which increase difficulty in longitudinal and lateral chromatic aberration correction respectively. In this study, the two chromatic aberrations are analyzed quantitatively based on an AOTF optical model and a novel catadioptric dual-path configuration is proposed to correct both the chromatic aberrations. The test results exhibit effectiveness of the optical configuration for this type of AOTF-based imaging spectrometer.

源语言英语
页(从-至)23809-23825
页数17
期刊Optics Express
25
20
DOI
出版状态已出版 - 2 10月 2017

学术指纹

探究 'Chromatic aberrations correction for imaging spectrometer based on acousto-optic tunable filter with two transducers' 的科研主题。它们共同构成独一无二的学术指纹。