摘要
In electrical capacitance tomography (ECT), the excitation frequency, excitation voltage, and the number of excitation electrode(s) can be chosen. The aim of this paper is to investigate their effect on ECT sensors with different diameter to improve the image quality and stability. Experimental studies have been carried out to understand the effect. To evaluate the experimental results objectively, the normalization models of capacitances are also examined. Using the linear back projection algorithm, images are reconstructed for six typical permittivity distributions, and the concentration is calculated based on the reconstructed images. Preliminary conclusions are given based on the experimental results.
| 源语言 | 英语 |
|---|---|
| 文章编号 | 6746647 |
| 页(从-至) | 2240-2251 |
| 页数 | 12 |
| 期刊 | IEEE Sensors Journal |
| 卷 | 14 |
| 期 | 7 |
| DOI | |
| 出版状态 | 已出版 - 7月 2014 |
| 已对外发布 | 是 |
指纹
探究 'Characterization of electrical capacitance tomography sensors with different diameter' 的科研主题。它们共同构成独一无二的指纹。引用此