TY - JOUR
T1 - Biomechanical effects of cervical arthroplasty with U-shaped disc implant on segmental range of motion and loading of surrounding soft tissue
AU - Mo, Zhong Jun
AU - Zhao, Yan Bin
AU - Wang, Li Zhen
AU - Sun, Yu
AU - Zhang, Ming
AU - Fan, Yu Bo
PY - 2014/3
Y1 - 2014/3
N2 - Purpose: Various design concepts have been adopted in cervical disc prostheses, including sliding articulation and standalone configuration. This study aimed to evaluate the biomechanical effects of the standalone U-shaped configuration on the cervical spine. Methods: Based on an intact finite element model of C3-C7, a standalone U-shaped implant (DCI) was installed at C5-C6 and compared with a sliding articulation design (Prodisc-C) and an anterior fusion system. The range of motion (ROM), adjacent intradiscal pressure (IDP) and capsular ligament strain were calculated under different spinal motions. Results: Compared to the intact configuration, the ROM at C5-C6 was reduced by 90 % after fusion, but increased by 70 % in the Prodisc-C model, while the maximum percentage change in the DCI model was 30 % decrease. At the adjacent segments, up to 32 % increase in ROM happened after fusion, while up to 34 % decrease occurred in Prodisc-C model and 17 % decrease in DCI model. The IDP increased by 11.6 % after fusion, but decreased by 5.6 and 6.3 % in the DCI and Prodisc-C model, respectively. The capsular ligament strain increased by 147 % in Prodisc-C and by 13 % in the DCI model. The DCI implant exhibited a high stress distribution. Conclusions: Spinal fusion resulted in compensatory increase of ROM at the adjacent sites, thereby elevating the IDP. Prodisc-C resulted in hyper-mobility at the operative site that led to an increase of ligament force and strain. The U-shaped implant could maintain the spinal kinematics and impose minimum influence on the adjacent soft tissues, despite the standalone configuration encountering the disadvantages of high stress distribution.
AB - Purpose: Various design concepts have been adopted in cervical disc prostheses, including sliding articulation and standalone configuration. This study aimed to evaluate the biomechanical effects of the standalone U-shaped configuration on the cervical spine. Methods: Based on an intact finite element model of C3-C7, a standalone U-shaped implant (DCI) was installed at C5-C6 and compared with a sliding articulation design (Prodisc-C) and an anterior fusion system. The range of motion (ROM), adjacent intradiscal pressure (IDP) and capsular ligament strain were calculated under different spinal motions. Results: Compared to the intact configuration, the ROM at C5-C6 was reduced by 90 % after fusion, but increased by 70 % in the Prodisc-C model, while the maximum percentage change in the DCI model was 30 % decrease. At the adjacent segments, up to 32 % increase in ROM happened after fusion, while up to 34 % decrease occurred in Prodisc-C model and 17 % decrease in DCI model. The IDP increased by 11.6 % after fusion, but decreased by 5.6 and 6.3 % in the DCI and Prodisc-C model, respectively. The capsular ligament strain increased by 147 % in Prodisc-C and by 13 % in the DCI model. The DCI implant exhibited a high stress distribution. Conclusions: Spinal fusion resulted in compensatory increase of ROM at the adjacent sites, thereby elevating the IDP. Prodisc-C resulted in hyper-mobility at the operative site that led to an increase of ligament force and strain. The U-shaped implant could maintain the spinal kinematics and impose minimum influence on the adjacent soft tissues, despite the standalone configuration encountering the disadvantages of high stress distribution.
KW - Cervical disc arthroplasty
KW - Finite element method
KW - Intradiscal pressure
KW - Ligament strain
KW - Range of motion
KW - U-shaped implant
UR - https://www.scopus.com/pages/publications/84896313731
U2 - 10.1007/s00586-013-3070-4
DO - 10.1007/s00586-013-3070-4
M3 - 文章
C2 - 24154828
AN - SCOPUS:84896313731
SN - 0940-6719
VL - 23
SP - 613
EP - 621
JO - European Spine Journal
JF - European Spine Journal
IS - 3
ER -