跳到主要导航 跳到搜索 跳到主要内容

Biomechanical comparison of implant inclinations and load times with the all-on-4 treatment concept: a three-dimensional finite element analysis

  • Ting Liu
  • , Zhixiang Mu
  • , Ti Yu
  • , Chao Wang
  • , Yuanding Huang*
  • *此作品的通讯作者
  • Chongqing Medical University
  • Chongqing Municipal Key Laboratory of Oral Biomedical Engineering of Higher Education
  • Chongqing Key Laboratory of Oral Diseases and Biomedical Sciences

科研成果: 期刊稿件文章同行评审

摘要

The purpose of this study was to compare the effects of implant inclinations and load times on stress distributions in the peri-implant bone based on immediate- and delayed-loading models. Four 3D FEA models with different inclination angle of the posterior implants (0°, 15°, 30°, 45°) were constructed. A static load of 150 N in the multivectoral direction was applied unilaterally to the cantilever region. The stress distributions in the peri-implant bone were evaluated before and after osseointegration. The principal tensile stress (σmax), mean principal tensile stress (σmax), principal compressive stress (σmin) and mean principal compressive stress (σmin) of the bone and micromotion at the contact interface between the bone and implants were calculated. In all the models, peak principal stresses occurred in the bone surrounding the left tilted implant. The highest σmax and σmin were all observed in the 0° model for both immediate- and delayed-loading models. And the 0° and 15° models showed higher σmax and σmin values. The 0°models showed the largest micromotion. The observed stress distribution was better in the 30° and 45° models than in the 0° and 15° models.

源语言英语
页(从-至)585-594
页数10
期刊Computer Methods in Biomechanics and Biomedical Engineering
22
6
DOI
出版状态已出版 - 26 4月 2019
已对外发布

指纹

探究 'Biomechanical comparison of implant inclinations and load times with the all-on-4 treatment concept: a three-dimensional finite element analysis' 的科研主题。它们共同构成独一无二的指纹。

引用此