TY - JOUR
T1 - Assessment of labially impacted canines traction mode with clear aligners vs. fixed appliance
T2 - A comparative study based on 3D finite element analysis
AU - Xia, Qian
AU - He, Yao
AU - Jia, Lurong
AU - Wang, Chunjuan
AU - Wang, Weixu
AU - Wang, Chao
AU - Song, Jinlin
AU - Fan, Yubo
N1 - Publisher Copyright:
Copyright © 2022 Xia, He, Jia, Wang, Wang, Wang, Song and Fan.
PY - 2022/10/5
Y1 - 2022/10/5
N2 - Purpose: The objective of this study was to evaluate and compare the biomechanical differences between clear aligner and fixed appliance in the traction of labially impacted canines based on 3D finite element analysis. Methods: A series of patient-oriented finite element models were constructed, including a maxillary dentition with a right labially canine, maxilla, periodontal ligaments, traction attachments, and clear aligners. The two most common clinical scenarios were investigated: Scenario A: impacted canine (distal) and Scenario B: impacted canine (mesial). For each clinical scenario, three traction models with clear aligners and one fixed appliance model were established. Results: In all four models, the impacted canines exhibited similar initial displacement tendencies of mesially rotated in Scenario A and distally rotated in Scenario B, and with small differences in periodontal ligament stress magnitude. However, the sum of the periodontal ligament stresses of the anchorage teeth in the clear aligner mode was in the range of 56.28–76.21 kPa and in the fixed appliance mode was in the range of 6.61–7.22 kPa. The maximum value of initial displacement of the anchorage teeth in the clear aligner mode was in the range of 13.71–19.72 μm, while in the fixed appliance mode was 3.10–3.92 μm. Conclusion: For impacted canines, clear aligner mode and fixed appliance mode have little difference in biomechanical effect. However, the anchorage teeth in the clear aligner mode endure higher stress and show a more pronounced displacement tendency. In addition, the biomechanical effects of different clear aligner traction models are various but not obvious.
AB - Purpose: The objective of this study was to evaluate and compare the biomechanical differences between clear aligner and fixed appliance in the traction of labially impacted canines based on 3D finite element analysis. Methods: A series of patient-oriented finite element models were constructed, including a maxillary dentition with a right labially canine, maxilla, periodontal ligaments, traction attachments, and clear aligners. The two most common clinical scenarios were investigated: Scenario A: impacted canine (distal) and Scenario B: impacted canine (mesial). For each clinical scenario, three traction models with clear aligners and one fixed appliance model were established. Results: In all four models, the impacted canines exhibited similar initial displacement tendencies of mesially rotated in Scenario A and distally rotated in Scenario B, and with small differences in periodontal ligament stress magnitude. However, the sum of the periodontal ligament stresses of the anchorage teeth in the clear aligner mode was in the range of 56.28–76.21 kPa and in the fixed appliance mode was in the range of 6.61–7.22 kPa. The maximum value of initial displacement of the anchorage teeth in the clear aligner mode was in the range of 13.71–19.72 μm, while in the fixed appliance mode was 3.10–3.92 μm. Conclusion: For impacted canines, clear aligner mode and fixed appliance mode have little difference in biomechanical effect. However, the anchorage teeth in the clear aligner mode endure higher stress and show a more pronounced displacement tendency. In addition, the biomechanical effects of different clear aligner traction models are various but not obvious.
KW - 3D printed
KW - biomechanics
KW - clear aligner
KW - finite element analysis
KW - labially impacted canine
UR - https://www.scopus.com/pages/publications/85140226839
U2 - 10.3389/fbioe.2022.1004223
DO - 10.3389/fbioe.2022.1004223
M3 - 文章
AN - SCOPUS:85140226839
SN - 2296-4185
VL - 10
JO - Frontiers in Bioengineering and Biotechnology
JF - Frontiers in Bioengineering and Biotechnology
M1 - 1004223
ER -