跳到主要导航 跳到搜索 跳到主要内容

Accuracy of cortical bone trajectory screw fixation guided by spinous process clamp guide in lumbosacral vertebrae: A cadaver study

  • Xi Nuo Zhang
  • , Li Jin Zhou*
  • , Qing Jun Su
  • , Li Guan
  • , Dong Yue Li
  • , Bao Qing Pei
  • , Ai Xing Pan
  • , Hong Hao Yang
  • , Hong Tao Ding
  • , Yu Zeng Liu*
  • , Yong Hai*
  • *此作品的通讯作者
  • Capital Medical University

科研成果: 期刊稿件文章同行评审

摘要

Background: The purpose of this study was to access the accuracy of cortical bone trajectory screw placement guided by spinous process clamp (SPC). Methods: Eight formalin-treated cadaveric lumbar specimens with T12-S1 were used. A total of 96 screws were implanted in eight lumbar specimens. Results: In the freehand (FH) group, clinically acceptable placement (grade A and B) was 40 screws (83.3%), meanwhile 44 screws (91.7%) in the SPC guide group (p = 0.217). The grade A screws in the SPC guide group were much more than that in the FH group (n = 40 vs. n = 31, p = 0.036). The misplacement screws (grade C, D, and E) and proximal facet joint violation (FJV) in the SPC group was comparable to the FH group. Conclusions: This cadaveric study demonstrate that implanting CBT screws guided by SPC guide was more accuracy and reduces severe deviations in important directions.

源语言英语
文章编号e2484
期刊International Journal of Medical Robotics and Computer Assisted Surgery
19
2
DOI
出版状态已出版 - 4月 2023

指纹

探究 'Accuracy of cortical bone trajectory screw fixation guided by spinous process clamp guide in lumbosacral vertebrae: A cadaver study' 的科研主题。它们共同构成独一无二的指纹。