A systematic study of independently-tuned room-specific PBS beam model in a beam-matched multiroom proton therapy system

  • Yu Hua Huang
  • , Chunfeng Fang
  • , Tao Yang
  • , Lin Cao
  • , Gaolong Zhang
  • , Baolin Qu
  • , Yihang Zhang
  • , Zishen Wang
  • , Shouping Xu*
  • *此作品的通讯作者
  • General Hospital of People's Liberation Army
  • Hong Kong Polytechnic University
  • Hebei Yizhou Tumor Hospital
  • Beihang University

科研成果: 期刊稿件文章同行评审

摘要

Background: In the existing application of beam-matched multiroom proton therapy system, the model based on the commissioning data from the leading treatment room was used as the shared model. The purpose of this study is to investigate the ability of independently-tuned room-specific beam models of beam-matched gantries to reproduce the agreement between gantries’ performance when considering the errors introduced by the modeling process. Methods: Raw measurements of two gantries’ dosimetric characteristics were quantitatively compared to ensure their agreement after initially beam-matched. Two gantries’ beam model parameters, as well as the model-based computed dosimetric characteristics, were analyzed to study the introduced errors and gantries’ post-modeling consistency. We forced two gantries to share the same beam model. The model-sharing patient-specific quality assurance (QA) tasks were retrospectively performed with 36 cancer patients to study the clinical impact of beam model discrepancies. Results: Intra-gantry comparisons demonstrate that the modeling process introduced the errors to a certain extent indeed, which made the model-based reproduced results deviate from the raw measurements. Among them, the deviation introduced to the IDD curves was generally larger than that to the beam spots during modeling. Cross-gantry comparisons show that, from the beam model perspective, the introduced deviations deteriorated the high agreement of the dosimetric characteristics originally shown between two beam-matched gantries, but the cross-gantry discrepancy was still within the clinically acceptable tolerance. In model-sharing patient-specific QA, for the particular gantry, the beam model usage for intensity-modulated proton therapy (IMPT) QA plan generation had no significant effect on the actual delivering performance. All reached a high level of 95.0% passing rate with a 3 mm/3% criterion. Conclusions: It was preliminary recognized that among beam-matched gantries, the independently-tuned room-specific beam model from any gantry is reasonable to be chosen as the shared beam model without affecting the treatment efficacy.

源语言英语
文章编号206
期刊Radiation Oncology
16
1
DOI
出版状态已出版 - 12月 2021

联合国可持续发展目标

此成果有助于实现下列可持续发展目标:

  1. 可持续发展目标 3 - 良好健康与福祉
    可持续发展目标 3 良好健康与福祉

学术指纹

探究 'A systematic study of independently-tuned room-specific PBS beam model in a beam-matched multiroom proton therapy system' 的科研主题。它们共同构成独一无二的学术指纹。

引用此