跳到主要导航 跳到搜索 跳到主要内容

A quartz crystal microbalance dew point sensor without frequency measurement

  • Guohua Wang*
  • , Weishuo Zhang
  • , Shuo Wang
  • , Jinglin Sun
  • *此作品的通讯作者
  • Beihang University

科研成果: 期刊稿件文章同行评审

摘要

This work deals with the design of a dew point sensor based on Quartz Crystal Microbalance (QCM) without measuring the frequency. This idea is inspired by the fact that the Colpitts oscillation circuit will stop oscillating when the QCM works in the liquid media. The quartz crystal and the electrode are designed through the finite element simulation and the stop oscillating experiment is conducted to verify the sensibility. Moreover, the measurement result is calibrated to approach the true value. At last a series of dew points at the same temperature is measured with the designed sensor. Results show that the designed dew point sensor is able to detect the dew point with the proper accuracy.

源语言英语
期刊论文编号115002
期刊Review of Scientific Instruments
85
11
DOI
出版状态已出版 - 1 11月 2014

学术指纹

探究 'A quartz crystal microbalance dew point sensor without frequency measurement' 的科研主题。它们共同构成独一无二的学术指纹。

引用此