T1 - A fully automatic 2D ultrasound probe calibration method using a simple phantom
AU - Wang, Junchen
AU - Wang, Tianmiao
AU - Hu, Lei
AU - Liao, Hongen
AU - Luan, Kuan
AU - Sakuma, Ichiro
PY - 2011
Y1 - 2011
N2 - Probe calibration should be performed for free-hand tracking and ultrasound (US) fusion with a position sensor-attached US probe in US guided interventions. Difficulties in locating the probe imaging plane and adjusting the alignment of the probe with the calibration phantom call for more complicated and precise phantoms. However, the manufacture of such phantoms is time-consuming and expensive. Furthermore, the segmentation of the phantom features in US images cannot be reliably automated, which requires manual intervention. This paper presents a fully automatic 2D probe calibration method using a rather simple phantom. By detecting and analyzing the temporal signal arising from the ultrasound artifacts when sweeping the 2D probe over the phantom, the imaging plane of the probe and the image features can be reliably identified. Then the features are fed to a nonlinear optimizer to estimate the calibration parameters. The whole procedure including the setup can be done within 2 minutes with the fiducial registration error (FRE) of 0.16 mm. A US-guided biopsy experiment has also been conducted which shows the target registration error (TRE) of less than 1 mm.
AB - Probe calibration should be performed for free-hand tracking and ultrasound (US) fusion with a position sensor-attached US probe in US guided interventions. Difficulties in locating the probe imaging plane and adjusting the alignment of the probe with the calibration phantom call for more complicated and precise phantoms. However, the manufacture of such phantoms is time-consuming and expensive. Furthermore, the segmentation of the phantom features in US images cannot be reliably automated, which requires manual intervention. This paper presents a fully automatic 2D probe calibration method using a rather simple phantom. By detecting and analyzing the temporal signal arising from the ultrasound artifacts when sweeping the 2D probe over the phantom, the imaging plane of the probe and the image features can be reliably identified. Then the features are fed to a nonlinear optimizer to estimate the calibration parameters. The whole procedure including the setup can be done within 2 minutes with the fiducial registration error (FRE) of 0.16 mm. A US-guided biopsy experiment has also been conducted which shows the target registration error (TRE) of less than 1 mm.
UR - https://www.scopus.com/pages/publications/84862937065
U2 - 10.1109/BMEI.2011.6098408
DO - 10.1109/BMEI.2011.6098408
M3 - 会议稿件
AN - SCOPUS:84862937065
SN - 9781424493524
T3 - Proceedings - 2011 4th International Conference on Biomedical Engineering and Informatics, BMEI 2011
SP - 588
EP - 592
BT - Proceedings - 2011 4th International Conference on Biomedical Engineering and Informatics, BMEI 2011
T2 - 2011 4th International Conference on Biomedical Engineering and Informatics, BMEI 2011
Y2 - 15 October 2011 through 17 October 2011
ER -