跳到主要导航 跳到搜索 跳到主要内容

A design of a low-power open-side MPI scanner Using inverted spliced permanent magnets

  • Tao Zhu
  • , Jie He
  • , Zechen Wei
  • , Hui Hui*
  • , Xin Yang
  • , Jie Tian
  • *此作品的通讯作者

科研成果: 期刊稿件文章同行评审

摘要

Magnetic particle imaging (MPI) is a promising imaging modality with high sensitivity, high resolution, and no depth limitation. Scaled MPI scanners are developed to satisfy the demand for imaging larger objects. Considering the ultimate goal of whole-body scanning, an open-side MPI scanner is more suitable than a traditional closed-bore MPI scanner. A fully electrically driven open-side MPI scanner has been built, enabling rotating and translating field free line (FFL) in three dimensions previously. However, the high power required to run the system may become a challenge when trying to scale it up. Here, we present a design of an open-side MPI device using inverted spliced permanent magnets. The design eliminates the need for power suppliers to generate and move the FFL, reducing the cooling requirements of the system. Compared to the conventional solution based on permanent magnets, our solution can generate a more uniform FFL.

源语言英语
文章编号2303072
期刊International Journal on Magnetic Particle Imaging
9
DOI
出版状态已出版 - 2023