TY - JOUR
T1 - A 402 MHz and 1.73-VCE Resonance Regulating Rectifier With On-Chip Antennas for Bioimplants
AU - Liu, Guoao
AU - Hu, Yuanqi
N1 - Publisher Copyright:
© 2007-2012 IEEE.
PY - 2025
Y1 - 2025
N2 - In this paper, a wireless power transfer (WPT) system composed of a voltage-mode fully integrated resonance regulating rectifier (IR3) and an on-chip antenna running at 402 MHz has been designed for bioimplants in deep tissue. The proposed IR3, including a 200 pF decoupling capacitor, is implemented in a 0.22 mm2 active area in the 180-nm CMOS process. A charging duration based regulation compensation circuit offers a low ripple factor of 0.3% at a 1.8 V output voltage and a high voltage conversion efficiency (VCE) of 1.73 to overcome the low inductive coupling coefficient (under 0.01) due to the deep implant scenario. And a clock gating VCDL-based on-&-off delay compensation scheme is proposed to compensate for the phase error of the IR3. Performing rectification and regulation simultaneously in a single stage, the IR3 effectively enhances power conversion efficiency. The whole system achieves a power conversion efficiency (PCE) of 65% with a 1.5 mW load. In addition, digital control-based compensation circuits also improve its transient response performance, the 1% setting time is only 6.9 μs when the load changes from 65 μW to 1.5 mW.
AB - In this paper, a wireless power transfer (WPT) system composed of a voltage-mode fully integrated resonance regulating rectifier (IR3) and an on-chip antenna running at 402 MHz has been designed for bioimplants in deep tissue. The proposed IR3, including a 200 pF decoupling capacitor, is implemented in a 0.22 mm2 active area in the 180-nm CMOS process. A charging duration based regulation compensation circuit offers a low ripple factor of 0.3% at a 1.8 V output voltage and a high voltage conversion efficiency (VCE) of 1.73 to overcome the low inductive coupling coefficient (under 0.01) due to the deep implant scenario. And a clock gating VCDL-based on-&-off delay compensation scheme is proposed to compensate for the phase error of the IR3. Performing rectification and regulation simultaneously in a single stage, the IR3 effectively enhances power conversion efficiency. The whole system achieves a power conversion efficiency (PCE) of 65% with a 1.5 mW load. In addition, digital control-based compensation circuits also improve its transient response performance, the 1% setting time is only 6.9 μs when the load changes from 65 μW to 1.5 mW.
KW - Regulating rectifier
KW - mm-sized implant
KW - wireless power transfer (WPT)
UR - https://www.scopus.com/pages/publications/85214308600
U2 - 10.1109/TBCAS.2024.3523913
DO - 10.1109/TBCAS.2024.3523913
M3 - 文章
C2 - 40030802
AN - SCOPUS:85214308600
SN - 1932-4545
VL - 19
SP - 968
EP - 980
JO - IEEE Transactions on Biomedical Circuits and Systems
JF - IEEE Transactions on Biomedical Circuits and Systems
IS - 5
ER -