高吻合度股胫关节面人工膝关节假体的磨损

  • Wenyu Ding
  • , Shuqin Ma
  • , Xingchen Hou
  • , Yujing Yang
  • , Yan Sun
  • , Feng Zhao*
  • , Yubo Fan
  • *此作品的通讯作者

科研成果: 期刊稿件文章同行评审

摘要

Objective To investigate the effect of the tibiofemoral joint surface with high conformity on total knee prosthesis motion and insert wear. Methods A type of knee prosthesis with a high coronal conformity was designed and manufactured, and its insert wear was studied by in vitro wear testing combined with finite element simulation. ResultsThe mass wear rates calculated by the in vitro wear testing and finite element simulation were (14.29±3.19) mg/MC and 14.67 mg/MC (MC, million cycle), respectively. After five MCs, the areas of the insert wear obtained by both the methods were basically consistent, and severe wear was found to occur in the middle part of the insert. ConclusionsThe result of the finite element simulation coincided with that of the in vitro testing. The wear rate of the tibiofemoral joint surface of the total knee prosthesis with a high coronal conformity was relatively low. The research outcomes have practical significance for the evaluation and clinical application of the total knee prosthesis with a high conformity.

投稿的翻译标题Wear Simulation of Tibiofemoral Joint Surface of Total Knee Prosthesis with High Conformity
源语言繁体中文
页(从-至)193-199
页数7
期刊Yiyong Shengwu Lixue/Journal of Medical Biomechanics
33
3
DOI
出版状态已出版 - 1 6月 2018

关键词

  • Finite element analysis
  • High conformity
  • In vitro wear testing
  • Total knee replacement