摘要
Objective To investigate the effect of the tibiofemoral joint surface with high conformity on total knee prosthesis motion and insert wear. Methods A type of knee prosthesis with a high coronal conformity was designed and manufactured, and its insert wear was studied by in vitro wear testing combined with finite element simulation. ResultsThe mass wear rates calculated by the in vitro wear testing and finite element simulation were (14.29±3.19) mg/MC and 14.67 mg/MC (MC, million cycle), respectively. After five MCs, the areas of the insert wear obtained by both the methods were basically consistent, and severe wear was found to occur in the middle part of the insert. ConclusionsThe result of the finite element simulation coincided with that of the in vitro testing. The wear rate of the tibiofemoral joint surface of the total knee prosthesis with a high coronal conformity was relatively low. The research outcomes have practical significance for the evaluation and clinical application of the total knee prosthesis with a high conformity.
| 投稿的翻译标题 | Wear Simulation of Tibiofemoral Joint Surface of Total Knee Prosthesis with High Conformity |
|---|---|
| 源语言 | 繁体中文 |
| 页(从-至) | 193-199 |
| 页数 | 7 |
| 期刊 | Yiyong Shengwu Lixue/Journal of Medical Biomechanics |
| 卷 | 33 |
| 期 | 3 |
| DOI | |
| 出版状态 | 已出版 - 1 6月 2018 |
关键词
- Finite element analysis
- High conformity
- In vitro wear testing
- Total knee replacement
指纹
探究 '高吻合度股胫关节面人工膝关节假体的磨损' 的科研主题。它们共同构成独一无二的指纹。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver