摘要
With the development of semiconductor photon detector technology, a novel signal readout device named multi pixel photon counter (MPPC) has gradually been developed and applied in radiation detection. [Purpose] This study aims to verify the performance of the MPPC and compare it with photomultiplier tube (PMT) in linearity and energy resolution. [Methods] First of all, scintillation detectors with MPPC coupling with scintillation crystals such as thallium-doped sodium iodide (NaI(Tl)), CsI, and Bi 4 Ge 3 O 12 (BGO) crystals was constructed to test three characteristic γ-ray energy peaks of 137 Cs and 60 Co. Then the same tests were done with PMT (Hamamatsu H7195) coupling with these crystals as the parallel experiments. Simultaneously, the output signal of the MPPC was collected by using a waveform digitizer (CAEN DT5730) and a self-made pre-amplifier. [Results] The results show that the MPPC has a slightly worse energy resolution compared to that of PMT, but reserves a similar energy linearity. [Conclusion] This MPPC could be applied in circumstances such as the space detection and medical applications instead of PMT.
| 投稿的翻译标题 | Energy resolution performance testing for mulit-pixel photon counter |
|---|---|
| 源语言 | 繁体中文 |
| 期刊论文编号 | 090403 |
| 期刊 | He Jishu/Nuclear Techniques |
| 卷 | 41 |
| 期 | 9 |
| DOI | |
| 出版状态 | 已出版 - 10 9月 2018 |
关键词
- Energy resolution
- Linearity
- Multi-pixel photon counter
- Scintillation crystal
- Waveform digitizer
学术指纹
探究 '多像素光子计数器的能量分辨性能测试' 的科研主题。它们共同构成独一无二的学术指纹。引用此