摘要
Steady and unsteady numerical simulations were carried out to systematically research the hemodynamic performance and hemocompatibility of four groups of centrifugal blood pumps with different blade tip clearance at 173mmHg. The gaps studied were 5.0mm, 2.0mm, 1.0mm and 0.2mm, respectively, the distribution of parameters and the changes of normalized index of hemolysis with the size of blade tip clearance were researched. The results showed that the origin pump had good hemocompatibility at 3.0L/min against a pressure head of 173mmHg; with the decrease of blade tip clearance, the max stress and the normalized index of hemolysis showed a single peak and the minimum normalized index of hemolysis 0.0005g/100L was obtained in the 1. 0mm clearance structure, which reduced the risk of thrombosis in the gap between the outer impeller walls and the lower housing cavity walls; compared with the results of the unsteady simulation: steady results were often too small to predict hemolysis, the transient normalized index of hemolysis changed periodically with the impeller rotation, and the difference between its mean and the steady results was within 5%.
| 投稿的翻译标题 | Numerical research on effects of blade tip clearances on the hemocompatibility of artificial heart pump |
|---|---|
| 源语言 | 繁体中文 |
| 页(从-至) | 1304-1314 |
| 页数 | 11 |
| 期刊 | Hangkong Dongli Xuebao/Journal of Aerospace Power |
| 卷 | 36 |
| 期 | 6 |
| DOI | |
| 出版状态 | 已出版 - 6月 2021 |
关键词
- Blade tip clearance
- Centrifugal pump
- Hemolysis
- Thrombosis
- Tip leakage vortex