叶顶间隙对人工心脏泵血液相容性影响的数值研究

  • Beihang University
  • Collaborative Innovation Center of Advanced Aero-Engine
  • Chinese Academy of Medical Sciences

科研成果: 期刊稿件文章同行评审

摘要

Steady and unsteady numerical simulations were carried out to systematically research the hemodynamic performance and hemocompatibility of four groups of centrifugal blood pumps with different blade tip clearance at 173mmHg. The gaps studied were 5.0mm, 2.0mm, 1.0mm and 0.2mm, respectively, the distribution of parameters and the changes of normalized index of hemolysis with the size of blade tip clearance were researched. The results showed that the origin pump had good hemocompatibility at 3.0L/min against a pressure head of 173mmHg; with the decrease of blade tip clearance, the max stress and the normalized index of hemolysis showed a single peak and the minimum normalized index of hemolysis 0.0005g/100L was obtained in the 1. 0mm clearance structure, which reduced the risk of thrombosis in the gap between the outer impeller walls and the lower housing cavity walls; compared with the results of the unsteady simulation: steady results were often too small to predict hemolysis, the transient normalized index of hemolysis changed periodically with the impeller rotation, and the difference between its mean and the steady results was within 5%.

投稿的翻译标题Numerical research on effects of blade tip clearances on the hemocompatibility of artificial heart pump
源语言繁体中文
页(从-至)1304-1314
页数11
期刊Hangkong Dongli Xuebao/Journal of Aerospace Power
36
6
DOI
出版状态已出版 - 6月 2021

关键词

  • Blade tip clearance
  • Centrifugal pump
  • Hemolysis
  • Thrombosis
  • Tip leakage vortex

指纹

探究 '叶顶间隙对人工心脏泵血液相容性影响的数值研究' 的科研主题。它们共同构成独一无二的指纹。

引用此