Abstract
To improve imaging quality of X-ray digital radiography system, an X-ray flat-panel-detector-based digital radiography system was investigated. Four factors of flat panel detector which degrade image quality, such as random noise, offset, response nonuniformity and defective pixels, were introduced. How they occur and influence imaging quality was analyzed. Corresponding methods for image calibration were put forward to reduce and eliminate the influences on image quality. Experimental results indicate that, with the image calibration methods, the influences of above mentioned factors can be removed effectively and the quality of image from flat panel detector can be improved greatly, which has founded a good foundation for subsquent image identification.
| Original language | English |
|---|---|
| Pages (from-to) | 698-701 |
| Number of pages | 4 |
| Journal | Beijing Hangkong Hangtian Daxue Xuebao/Journal of Beijing University of Aeronautics and Astronautics |
| Volume | 30 |
| Issue number | 8 |
| State | Published - Aug 2004 |
Keywords
- Flat panel detector
- Image calibration
- Image processing
- Nondestructive test
Fingerprint
Dive into the research topics of 'X-ray flat-panel-detector-based digital radiography and its image calibration'. Together they form a unique fingerprint.Cite this