The Research of global correction for c-arm x-ray image based on pin-hole model

  • Xiaojun Zhou*
  • , Cai Meng
  • , Limin Fan
  • , Shaojie Lv
  • *Corresponding author for this work

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

Abstract

There are extensive studies had been carried out for the correction of C arm X-ray image distortion over the world. The global method specializes in analysing the establishment of ideal points during the process of image correction, and this method was adopted to be used in the study of this article. Compare with the traditional method of establishing ideal points, Global Method suggests creating the ideal points by using the pin-hole model, which could improve the precision of image correction. Meanwhile, it analyses the effects on image correction caused by deviations of calibration phantoms installation. It could be an important foundation for further improvement on surgical navigation.

Original languageEnglish
Title of host publicationProceedings of the 2009 2nd International Conference on Biomedical Engineering and Informatics, BMEI 2009
DOIs
StatePublished - 2009
Event2009 2nd International Conference on Biomedical Engineering and Informatics, BMEI 2009 - Tianjin, China
Duration: 17 Oct 200919 Oct 2009

Publication series

NameProceedings of the 2009 2nd International Conference on Biomedical Engineering and Informatics, BMEI 2009

Conference

Conference2009 2nd International Conference on Biomedical Engineering and Informatics, BMEI 2009
Country/TerritoryChina
CityTianjin
Period17/10/0919/10/09

Keywords

  • Fixing error
  • Global-corrcection method
  • Ideal points' establishment
  • Pin-hole model

Fingerprint

Dive into the research topics of 'The Research of global correction for c-arm x-ray image based on pin-hole model'. Together they form a unique fingerprint.