Skip to main navigation Skip to search Skip to main content

The blood heat exchanger in intra-arterial selective cooling infusion for acute ischemic stroke: A computational fluid-thermodynamics performance, experimental assessment and evaluation on the brain temperature

  • Miaowen Jiang
  • , Yuan Gao
  • , Chuanjie Wu
  • , Longfei Wu
  • , Shoucheng Tang
  • , Zhichen Yin
  • , Ang Li
  • , Kun Wang
  • , Shiqiang Zheng
  • , Hangil Lee
  • , Yuchuan Ding
  • , Ming Li*
  • , Xunming Ji*
  • *Corresponding author for this work
  • Beihang University
  • Capital Medical University
  • CAS - Technical Institute of Physics and Chemistry
  • Columbia University
  • Wayne State University
  • BUAA-CCMU Advanced Innovation Center for Big Data-based Precision Medicine

Research output: Contribution to journalArticlepeer-review

Abstract

Intra-arterial selective cooling infusion with the autologous blood (IA-SCAI) is a promising therapeutic hypothermia induction method for conferring neuroprotection to acute ischemic stroke (AIS) patients. The blood heat exchanger (BHE) plays a crucial role in IA-SCAI's cooling capacity. However, there are no BHEs currently available that are specifically designed for the IA-SCAI, which requires a low blood flow to be compatible with cerebral hemodynamics. In an effort to develop a BHE for AIS patients, a prototype of a commercial BHE, Medtronic MYOtherm XP®, was mathematically modeled; specifically, computational fluid dynamics (CFD) was used to analyze its hemo- and thermo-dynamic characteristics under low blood flow including temperature distribution, velocity field and shear stress. Our numerical model predicted the hemolysis index to be 0.0041%–0.0581% inside the BHE with blood flows rates of 10 ml min−1–50 ml min−1. The in vitro heat transfer experiment showed that the BHE efficiently cooled the simulated blood from the initial 37 °C–5.8 °C within 150 s by using cold water (200 ml·min-1, 0 °C). The cooled simulated blood was able to cool the simulated blood in the middle cerebral artery of an artificial circulating system from 37 °C to 16.8 °C–33.7 °C depending on the blood perfusion rate (10–50 ml/min). A biological heat transfer mathematical model showed that brain tissue could be cooled by 2 °C within the initial 1min of infusion. This study verified the feasibility of using a commercial BHE for IA-SCAI and provided insights into its cooling capacity for therapeutic hypothermia.

Original languageEnglish
Article number105497
JournalComputers in Biology and Medicine
Volume145
DOIs
StatePublished - Jun 2022

Keywords

  • Blood heat exchanger
  • Hemolysis
  • Numerical simulation
  • Stroke
  • Therapeutic hypothermia

Fingerprint

Dive into the research topics of 'The blood heat exchanger in intra-arterial selective cooling infusion for acute ischemic stroke: A computational fluid-thermodynamics performance, experimental assessment and evaluation on the brain temperature'. Together they form a unique fingerprint.

Cite this