Abstract
High-performance magnetic shielding systems require accurate shielding factor (SF) evaluation, yet testing ultra-high-performance systems (SF > 108) remains challenging due to the need for impractically large external excitation fields. We developed a testing system specifically designed for this task, based on a portable atomic magnetometer, achieving a sensitivity better than 1 f T/Hz1/2. By applying an order of 104 nT external field via uniform coils and extracting signals via noise spectrum analysis, we achieved the capability to test SFs up to the order of 1010 at a signal-to-noise ratio of 2. Testing of 10-layer magnetic shielding cylinders revealed a pronounced frequency dependence of the SF within the 1–100 Hz band. The radial SF ranges from 1.7 × 109 to 5.0 × 1010, while the axial SF varies between 4.8 × 107 and 1.3 × 1010, exhibiting the characteristic of an initial increase followed by a decrease with frequency. Furthermore, the phase-frequency characteristics, which are often overlooked in previous studies, were also tested. The results revealed pronounced hysteresis, which is of great significance for practical applications involving real-time magnetic field closed-loop systems.
| Original language | English |
|---|---|
| Article number | 367001 |
| Journal | Measurement Science and Technology |
| Volume | 37 |
| Issue number | 36 |
| DOIs |
|
| State | Published - Sep 2026 |
Keywords
- magnetic shield
- shielding factor (SF)
- spin-exchange relaxation-free (SERF) magnetometer
Fingerprint
Dive into the research topics of 'Testing of magnetic shielding system exceeding 1010 damping based on ultrasensitive atomic magnetometer'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver