Skip to main navigation Skip to search Skip to main content

Tensional analysis and suturing performance evaluation of the suture in corneal suturing

  • Yang Yang*
  • , Lu Wang
  • , Shijing Deng
  • *Corresponding author for this work
  • Beihang University
  • Beijing Institute of Ophthalmology

Research output: Contribution to journalArticlepeer-review

Abstract

According to the technical requirements for the corneal suturing in cornea transplantation, a finite element model for the interaction of the SMA (shape memory alloy) suture for corneal suturing and the soft tissue of the cornea was built to explore the effect of the flexible suture on the viscoelastic cornea. Moreover, the influence of the suture tension on the maximum displacement and stress of the corneal tissue was analyzed. Subsequently, the suturing performance evaluating method for the clinical surgery with minimal trauma was put forward. The proposed suturing performance evaluating method can offer the reference information for suturing operations, which can lay a good foundation for the further research on the self-tying suture.

Original languageEnglish
Pages (from-to)74-78
Number of pages5
JournalGaojishu Tongxin/Chinese High Technology Letters
Volume23
Issue number1
DOIs
StatePublished - Jan 2013

Keywords

  • Corneal stuturing
  • Evaluation
  • Finite element model
  • Suture
  • Tensional analysis

Fingerprint

Dive into the research topics of 'Tensional analysis and suturing performance evaluation of the suture in corneal suturing'. Together they form a unique fingerprint.