Summing coincidence correction for γ-ray measurements using the HPGe detector with a low background shielding system

  • L. C. He
  • , L. J. Diao
  • , B. H. Sun*
  • , L. H. Zhu
  • , J. W. Zhao
  • , M. Wang
  • , K. Wang
  • *Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

Abstract

A Monte Carlo method based on the GEANT4 toolkit has been developed to correct the full-energy peak (FEP) efficiencies of a high purity germanium (HPGe) detector equipped with a low background shielding system, and moreover evaluated using summing peaks in a numerical way. It is found that the FEP efficiencies of 60Co, 133Ba and 152Eu can be improved up to 18% by taking the calculated true summing coincidence factors (TSCFs) correction into account. Counts of summing coincidence γ peaks in the spectrum of 152Eu can be well reproduced using the corrected efficiency curve within an accuracy of 3%.

Keywords

  • GEANT4
  • HPGe detector
  • Low background shielding system
  • Summing coincidence