Skip to main navigation Skip to search Skip to main content

Static characteristics of a mN range thrust stand and uncertainty analysis of thrust measurements

  • Chen Bo Shi*
  • , Hai Bin Tang
  • , Xin Ai Zhang
  • , Chao Jin Qin
  • , Xi Wei Cao
  • *Corresponding author for this work
  • Beihang University

Research output: Contribution to journalArticlepeer-review

Abstract

In order to obtain thrust stand static characteristics including sensitivity, linearity, hysteresis, repetition, resolution and long-term stability, the thrust stand were statically calibrated by means of weight measurement. It is proven that the feasibility of electromagnetic force technique measuring mN range thrust. Effect factors were analyzed; thrust produced by a cold gas thruster was measured by the thrust stand as a validation test. The results show that uncertainty of the thrust is mainly ascribed to static characteristics. During measurement, pumps' vibration may cause change of force produced by the propellant pipe, leading to thrust stand zero drift output, which may also increase uncertainty.

Original languageEnglish
Pages (from-to)398-402
Number of pages5
JournalGuti Huojian Jishu/Journal of Solid Rocket Technology
Volume34
Issue number3
StatePublished - Jun 2011

Keywords

  • Static calibration
  • Static characteristics
  • Uncertainty
  • mN range thrust stand

Fingerprint

Dive into the research topics of 'Static characteristics of a mN range thrust stand and uncertainty analysis of thrust measurements'. Together they form a unique fingerprint.

Cite this