Shear stress and blood trauma under constant and pulse-modulated speed CF-VAD operations: CFD analysis of the HVAD

  • Zengsheng Chen
  • , Sofen K. Jena
  • , Guruprasad A. Giridharan
  • , Michael A. Sobieski
  • , Steven C. Koenig
  • , Mark S. Slaughter
  • , Bartley P. Griffith
  • , Zhongjun J. Wu*
  • *Corresponding author for this work
  • University of Maryland, Baltimore
  • University of Louisville
  • University of Maryland, College Park

Research output: Contribution to journalArticlepeer-review

Abstract

Modulation of pump speed has been proposed and implemented clinically to improve vascular pulsatility in continuous flow ventricular assist device patient. The flow dynamics of the HVAD with a promising asynchronous pump speed modulation and its potential risk for device-induced blood trauma was investigated numerically. The boundary conditions at the pump inlet and outlet were defined using the pressure waveforms adapted from the experimentally recorded ventricular and arterial pressure waveforms in a large animal ischemic heart failure (IHF) model supported by the HVAD operated at constant and modulated pump speeds. Shear stress fields and hemolysis indices were derived from the simulated flow fields. The overall features of the computationally generated flow waveforms at simulated constant and pulse-modulated speed operations matched with those of the experimentally recorded flow waveforms. The simulations showed that the shear stress field and hemolysis index vary throughout the cardiac cycle under the constant speed operation, and also as a function of modulation profile under modulated speed operation. The computational model did not demonstrate any differences in the time average hemolysis index between constant and modulated pump speed operations, thereby predicting pulse-modulated speed operation may help to restore vascular pulsatility without any further increased risk of blood trauma. [Figure not available: see fulltext.].

Original languageEnglish
Pages (from-to)807-818
Number of pages12
JournalMedical and Biological Engineering and Computing
Volume57
Issue number4
DOIs
StatePublished - 11 Apr 2019
Externally publishedYes

Keywords

  • Blood trauma
  • Continuous flow ventricular assist devices
  • Pulsatility
  • Shear stress
  • Speed pulse modulation

Fingerprint

Dive into the research topics of 'Shear stress and blood trauma under constant and pulse-modulated speed CF-VAD operations: CFD analysis of the HVAD'. Together they form a unique fingerprint.

Cite this