Sequential Scan-Based Single-Dimension Multi-Voxel System Matrix Calibration for Open-Sided Magnetic Particle Imaging

  • Jie He
  • , Haoran Zhang
  • , Yimeng Li
  • , Guanghui Li
  • , Siao Lei
  • , Zhumei Qian
  • , Fei Xiong
  • , Yuan Feng
  • , Tao Zhu
  • , Yu An*
  • , Jie Tian*
  • *Corresponding author for this work
  • Beihang University
  • CAS - Institute of Automation

Research output: Contribution to journal › Article › peer-review

Abstract

Open-sided magnetic particle imaging (OS-MPI) has garnered significant interest due to its potential for interventional applications. However, the system matrix calibration in OS-MPI using sequential scans is a time-consuming task and susceptible to the low signal-to-noise ratio (SNR) resulting from the small calibration sample size. These challenges have hindered the practical implementation of system matrix-based reconstruction for sequentially scanned OS-MPI. To address these issues, we propose a novel calibration method, named sequen- tial scan-based single-dimension multi-voxel calibration (SS-SDMVC), to efficiently obtain a high-SNR system matrix. This method was implemented in a cylindrical field of view (FOV), where a bar calibration sample parallel to the field-free line (FFL) was shifted along a fixed radial direction. A standard image reconstruction process was also introduced to verify the feasibility of SS-SDMVC. Through simulations, we analyzed the effects of noise levels and scanner imperfections on the SS-SDMVC-based reconstruction and demonstrated its robustness. In experiments, we compared the imaging performance of SS-SDMVC and the sequential scan-based traditional cubic-FOV SMC. The results showed that SS-SDMVC reduced the number of measurements by a factor of 210.94 and achieved higher reconstruction quality. Therefore, SS-SDMVC is expected to improve the reconstruction quality of human- scale or high-gradient FFL MPI scanners.

Original languageEnglish
Pages (from-to)3856-3868
Number of pages13
JournalIEEE Transactions on Medical Imaging
Volume43
Issue number11
DOIs
StatePublished - 2024

Keywords

  • Magnetic particle imaging
  • field-free line
  • open-sided
  • sequential scan
  • system matrix calibration