Polarization calibration with large apertures in full field of view for a full Stokes imaging polarimeter based on liquid-crystal variable retarders

Research output: Contribution to journalArticlepeer-review

Abstract

Currently, polarization calibration for full Stokes imaging polarimeters is limited by the apertures of the retarders. In this paper, an improved polarization calibration with large apertures in full field of view for full Stokes imaging polarimeters based on liquid-crystal variable retarders is proposed and investigated theoretically and experimentally. The experimental precision of polarization calibration is 1.7% for linear polarization states and 8.8% for circular ones for an imaging polarimeter witha 100 mmaperture and 10° field of view. The feasibility for full Stokes polarization image is also confirmed in experiment for identifying objects due to degree of polarization and degree of circular polarization images.

Original languageEnglish
Pages (from-to)1284-1292
Number of pages9
JournalApplied Optics
Volume52
Issue number6
DOIs
StatePublished - 20 Feb 2013