Abstract
Objective To study the flow field of a new vascular bioreactor with small diameter at different operating conditions. Method To investigate the flow field of a new vascular bioreactor with small diameter under three operating conditions by numerical simulation: outer-box rotating, inner-box rotating and both inner-box and outer-box rotating. A comparison of parameters such as flow velocity and shear stress was made. Results The culture medium in the bioreactor rotated with the inner-box and outer-box, and the velocity was uniformly distributed; the bioreactor could therefore provide a low shear stress environment that is nondestructive for the cells in the medium; the shear stress in bioreactor had a quadratic function relation with the rotation rate. Conclusions This bioreactor could provide a good culture environment for vascular tissue regeneration, and this study provided theoretical basis for the experimental research of bioreactors.
| Original language | English |
|---|---|
| Pages (from-to) | 352-356+368 |
| Journal | Yiyong Shengwu Lixue/Journal of Medical Biomechanics |
| Volume | 25 |
| Issue number | 5 |
| State | Published - Oct 2010 |
Keywords
- Flow velocity
- Numerical simulation
- Rotational flow
- Shear stress
- Vascular bioreactor
Fingerprint
Dive into the research topics of 'Numerical analysis on flow field in a vascular bioreactor'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver